Customer support
submenu
Welcome to UJP Medical Devices Online Services
Customer support for medical devices produced by company UJP PRAHA a.s. covers all medical devices (products) that the company places on the market. Description and information about each version is in Czech-English language version. According to the requirements of Regulation (EU) 2017/745 of the European Parliament the manufacturer guarantees the supplying of information in the appropriate language of the user. Contact sales department to provide this information.
TERAGIS RVS
Version
Date of release
Description of changes
1.1.32
10.05.2024
UDI implementation.
1.1-31
29.1.2018
Support for Cyrillic in imported DICOM files was added.
1.1-30
3.7.2017
MLC (Multi Leaf Collimator) support added. The position of the MLC leaves is transferred from the planning system using the DICOM protocol. The positions are transferred to the TERASIX simulator.
1.1-29
26.4.2017
The system was allowed to enter working and nonworking days. Earlier versions anticipated weekends as Saturdays and Sundays; this behavior can now be freely configured, e.g. set as nonworking days Friday + Saturday (Arab countries).
1.1-28
14.1.2016
The list of diagnoses has been updated and extended by the Russian language. The language version of the diagnosis is automatically switched according to the selected language version of TERAGIS.
PlanW TPS
Version
Date of release
Description of changes
3.0.2023
04.12.2023
Change of version designation due to hardware version limitation time. UDI was added.
3.0.2022 – 26.09.2022
26.09.2022
Change of version designation due to hardware version limitation time.
3.0.2020 – 21.12.2020
21.12.2020
Change of version designation due to hardware version limitation time.
3.0.2019 – 12.12.2019
12.12.2019
Superficial in-vivo dosimetry protocol is now available in the system.
New keyboard shortcuts have been implemented, especially into the Planning module.
3.0.2018 – 27.08.2018
27.08.2018
Digitizer – All buttons are reachable by lens cursor.
The display of individual parts of the shielding block is fine-tuned.
The refresh of displayed iso position is fixed.
The functionality of automatic selection of contour part that should be deleted is fixed.
Export of dose distribution locally to disk is fixed.
Decimal separators are dots.
The font size and font in the plan protocol footer are unified.
The maximum transmission of the backup jaw is 20%.
TPR value is dispalyed in an in-vivo dosimetry protocol.
When the RT plan is exported, it is checked. The result is recorded in a log file.
When the radiation type is changed for one field it does not affect the radiation type of other fields.
When dose distribution is exported, the DICOM tags 3004,000a a 300c,0006 are filled correctly.
The DD calculation option is active when DynamicArc field is added.
The display of patient position for 90° couch rotation is fixed.
CT slices entered in graphical format (scanned image) can be exported.
VOI frame functionality is fine-tuned.
The SOPinstanceUID is filled when RT structures delineated in phantom are exported.
DICOM log is deleted gradually – row by row.
The contour rendering using a specific procedure is fine-tuned.
Unification of keyboard shortcuts used in TPS individual modules with familiar shortcuts known from OS.
3.0.2017 – 16.11.2017
16.11.2017
Change of CE mark (new DNV number).
3.0.2017 – 6.10.2017
6.10.2017
Export of calculated dose distribution – change of the file name from PET.xxx to RTD.xxx
3.0.2017 – 27.3.2017
27.03.2017
Support for the import of dosimetric data in Sun Nuclear format.
3.0.2017 – 31.1.2017
31.01.2017
Name of field parameter was added into the printout of approved plan.
3.0.2016 – 29.3.2016
29.03.2016
Modification of the look of PlanW installation wizard.
3.0.2016 – 8.1.2016
08.01.2016
The opening a so-called portal plan on a simulator or CT simulator and enabling 2D and 3D position verification is fixed.